Recall archive / FDA device enforcement
Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope... — FDA record Z-2522-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope. Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012
- Reason
- The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
- Classification
- Class II
- Recalling firm
- Aomori Olympus Co., Ltd.
- Quantity
- 1,048 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2522-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | (1)K-001: Lot Numbers: 5835330 0XI, 0YI, 12I, 14I, 15I, 16I, 17I, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK, 31K, 32K. UDI: 4953170048562 (2) K-002 Lot Numbers: 5835430 0XI, 0YI, 12I, 13I, 14I, 16I, 17I, 27K, 2XK, 2YK, 2ZK 4953170048579 K-003 5835530 0XI, 0YI, 12I, 13I, 14I, 15I, 16I, 17I, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK UDI:4953170048586 (3)K-004 Lot Numbers: 5835630 16I, 17I, 26K UDI: 4953170048593 (4)K-011 Lot Numbers: N2485230 0XI, 15I, 17I UDI: 4953170214035 (5) K-012 Lot Numbers: N2485330 16I, 17I UDI: 4953170214042 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Sep 13, 2023
- Recall initiation date
- Jul 27, 2023
- Center classification date
- Sep 7, 2023