Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: ... — FDA record Z-2520-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 13, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version: N/A Product Description: The HeartMate 11 Volt Lithium Ion Standalone Backup Battery is a component of the HeartMate 3" System Controller and provides power to the System Controller. The HeartMate System Controller is an external device that manages the HeartMate Left Ventricular Assist System (LVAS), including alarms, system monitoring, and communication. It contains an internal 11V lithium-ion backup battery that provides at least 15 minutes of emergency power should the main system power source fail. The backup battery must be properly installed, charges automatically during normal use, and has a 36-month service life from date of manufacture. Component: Yes, HeartMate 3" System Controller
Reason
Due to 11 volt Backup Battery failures.
Classification
Class II
Recalling firm
Thoratec LLC
Quantity
12
Agency-reported status
Ongoing
Source record identifier
Z-2520-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Model Number: 106128 UDI/GTIN 05415067039880 Serial Numbers Distributed in US: See Attachment E for serial numbers distributed.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Canada, Germany, Japan, Netherlands, Poland, Spain.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 24, 2026
Recall initiation date
May 13, 2026
Center classification date
Jun 18, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.