Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The... — FDA record Z-2518-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 11, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.
Reason
Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.
Classification
Class II
Recalling firm
Merge Healthcare, Inc.
Quantity
84
Agency-reported status
Ongoing
Source record identifier
Z-2518-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySchiller Model Name: ARGUS PB-3000, Schiller Model #: 3.921002, Affected PB-3000 Serial Numbers: 7012.000035, 7012.000036, 7012.000037, 7012.000038, 7012.000041, 7012.000042, 7012.000044, 7012.000045, 7012.000046, 7012.000047, 7012.000048, 7012.000050, 7012.000051, 7012.000052, 7012.000053, 7012.000057, 7012.000058, 7012.000059, 7012.000060, 7012.000061, 7012.000062, 7012.000063, 7012.000064, 7012.000065, 7012.000066, 7012.000067, 7012.000068, 7012.000069, 7012.000071, 7012.000072, 7012.000073, 7012.000074, 7012.000075, 7012.000076, 7012.000078, 7012.000079, 7012.000080, 7012.000081, 7012.000082, 7012.000083, 7012.000085, 7012.000087, 7012.000088, 7012.000089, 7012.000090, 7012.000091, 7012.000092, 7012.000093, 7012.000094, 7012.000097, 7012.000098, 7012.000099, 7012.000100, 7012.000101, 7012.000102, 7012.000103, 7012.000104, 7012.000105, 7012.000106, 7012.000107, 7012.000108, 7012.000109, 7012.000110, 7012.000111, 7012.000112, 7012.000113, 7012.000114, 7012.000115, 7012.000116, 7012.000117, 7012.000118, 7012.000119, 7012.000120, 7012.000123, 7012.000124, 7012.000125, 7012.000126, 7012.000137, 7012.000138, 7012.000140, 7012.000141, 7012.000142, 7012.000157, 7012.000176

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution including in the states of Arizona, Arkansas, California, Florida, Louisiana, Mississippi, North Carolina, North Dakota, Oklahoma, Pennsylvania, Georgia, Illinois, Kentucky, Tennessee, Texas, Washington, and Wisconsin.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 10, 2025
Recall initiation date
Aug 11, 2025
Center classification date
Sep 4, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records