Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product D... — FDA record Z-2516-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 11, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire¿ Stimulation Lead (Model 4063) is designed to deliver stimulation to the hypoglossal nerve for the treatment of obstructive sleep apnea. The lead features a flexible, self-sizing cuff. Electrodes in the inner surface of the cuff deliver stimulation to the nerve. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Reason
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
Classification
Class II
Recalling firm
Inspire Medical Systems Inc.
Quantity
10 impacted units
Agency-reported status
Ongoing
Source record identifier
Z-2516-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Model 4063 (USA only lead); UDI-DI [4063] 10855728005793; Inspire part number [4063] 900-013-001; serial numbers D53393, D61104 Model 4063 (Global lead); UDI-DI [4063] 10810098650295, Inspire part number [4063] 900-013-003; serial numbers D89539, D89540, D89543, D89559, D89563, D89564, D89577, D8959

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the country of Hong Kong.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 24, 2026
Recall initiation date
May 11, 2026
Center classification date
Jun 18, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records