Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calcul... — FDA record Z-2509-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 20, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
Reason
Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
Classification
Class II
Recalling firm
Staar Surgical AG
Quantity
543
Agency-reported status
Ongoing
Source record identifier
Z-2509-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: UDI (01)00840311346288(0812)8.00 Software Version 8.00 The software error occurs if a customer generated an IOD for a toric implantation from April 13, 2026 to April 16, 2026 for multiple toric lens reservations for the same eye using the STAAR ICL Calculation Software.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Austria, Australia, Belgium, Canada, Switzerland, Czech Republic, Germany Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, India, Iran, Italy, Jordan, Korea, Kuwait, Kazakhstan, Morocco, Malaysia, Netherlands, Oman, Philippines, Poland, Saudi Arabia, Sweden, Singapore, Thailand, Turkey, Taiwan, Ukraine, USA, Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 24, 2026
Recall initiation date
Apr 20, 2026
Center classification date
Jun 17, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.