Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis im... — FDA record Z-2481-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 26, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty surgical procedure Item Number: 17-5220/01
Reason
A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery
Classification
Class II
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
6 units
Agency-reported status
Ongoing
Source record identifier
Z-2481-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04026575284658 Lot Codes: C010108, C151519

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

IN,GA, TX

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 6, 2023
Recall initiation date
Jul 26, 2023
Center classification date
Aug 29, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records