Recall archive / FDA device enforcement
PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS — FDA record Z-2479-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS
- Reason
- Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
- Classification
- Class II
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 92
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2479-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Basic-UDI Code: 0690103S004PAS000BC. Model/Serial Numbers: 20000ISM/10844549, 10844558, 10921747, 10700424, 10809849, 10809852, 10844564, 10864947, 10903936, 10903937, 10921752, 10921754, 10865580, 10921757, 10921758, 10921760, 10700439, 10776944, 10776966, 10776982, 10809844, 10844243, 10844251, 10844281, 10844292, 10844571, 10848756, 10864921, 10864926, 10864927, 10864942, 10864943, 10864945, 10864950, 10865585, 10865595, 10865603, 10865652, 10865653, 10865654, 10865656, 10865720, 10865723, 10865724, 10865730, 10865731, 10903944, 10903960, 10921729, 10921739, 10921922, 10921924, 10921925, 10921932, 10921941, 10844264, 10844284, 10844288, 10844581, 10844585, 10848729, 10848730, 10848739, 10848740, 10848744, 10848746, 10848747, 10848750, 10848753, 10864833, 10864838, 10864840, 10864844, 10864897, 10865584, 10865613, 10865616, 10865618, 10865649, 10865719, 10865722, 10865727, 10921733, 10921734, 10921918, 10951672, 10951679, 10951683, 10951685, 10951687, 10951691, 10865732, 10865735, 10903975, 10809836, 10848723, 10921930, 10844300, 10844302, 10844305, 10864959, 10864960, 10921926, 10921927, 10844551, 10844563, 10844566, 10853110, 10903977, 10921748, 10921750, 10922081, 10922089, 10922093, 10922096, 10922098, 10922102, 10922103, 10922106, 10922107, 10922108, 10922110, 10922111, 10922116, 10922118, 10922119, 10922121, 10922122, 10922126, 10922127, 10922129, 10922133, 10922138, 10922140, 10922142, 10951734, 10951737, 10951740, 10951741, 10951743, 10951744, 10951745, 10951747, 10951748, 10951749, 10951750, 10951751, 10951753, 10951754, 10951756, 10951757, 10951758, 10951760, 10951761, 10951763, 10969562, 10969572, 10969573, 10969576, 10969577, 10969579, 10969583, 10969587, 10969591, 10969592, 10969593, 10969595, 10969597, 10969598, 10969599, 10969601, 10969603, 10969606, 10969611, 10969616, 10969619, 10969620, 10969625, 10969629, 10848722, 10865736, 10904022, 10904024, 10904026, 10904028, 10904029, 10864924, 10864928, 10864931, 10864936, 10864939, 10865549, 10865716, 10921861. 20000IS/10875150, 10853171, 10853233, 10853196, 10853203, 10853208, 10853213, 10853222, 10853235, 10875121, 10875124, 10875126, 10875131, 10875137, 10875138, 10875143, 10875145, 10875146, 10875152, 10853206, 10875154, 10875123, 10875136, 10875142, 10875144 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
International distribution to the countries of Austria, Belgium, Switzerland, Germany, Cyprus, Israel, UK, Greece, Ireland, Italy, Luxembourg, Poland.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Sep 6, 2023
- Recall initiation date
- Jul 27, 2023
- Center classification date
- Aug 29, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.