Recall archive / FDA device enforcement
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANC... — FDA record Z-2464-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 14, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB; 3) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48O; 4) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48P; 5) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18F; 6) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18G; 7) CABG ACCESSORY PACK, MODEL/ITEM NUMBER HGCB98F; 8) CV SET UP PACK, MODEL/ITEM NUMBER HGCV05K; 9) OPEN HEART PACK B - COMPONENTS, MODEL/ITEM NUMBER HGOH99H; 10) STRADLE TABLE PACK - 301659, MODEL/ITEM NUMBER MHCV10AM; 11) MINOR VASCULAR PACK, MODEL/ITEM NUMBER MTVS73D; 12) AV GRAFT RCH 0246952, MODEL/ITEM NUMBER RCAV40C; 13) OPEN HEART PART 1 RCH 0247159, MODEL/ITEM NUMBER RCOH61G; 14) CARDIO 612968, MODEL/ITEM NUMBER TNCA08T; 15) CV B PACK, MODEL/ITEM NUMBER TNCB10AM; 16) A-V FISTULA - 206042, MODEL/ITEM NUMBER UDAV16AD; 17) CAROTID PACK - 206045, MODEL/ITEM NUMBER UDCP24AF; 18) FEM POP PACK - 206052, MODEL/ITEM NUMBER UDFP36AH; 19) CARDIOVASCULAR SUPPLY (PS 907285), MODEL/ITEM NUMBER UICD62AN; 20) SFCH PEDS CARDIAC SUPPLY PACK (PS131653), MODEL/ITEM NUMBER UICS88S; 21) SFCH ECMO PACK (PS 107116), MODEL/ITEM NUMBER UIPD03Q; 22) THORACIC SETUP (PS 035700), MODEL/ITEM NUMBER UITC53AL; 23) VASCULAR PACK, MODEL/ITEM NUMBER UIVA75J; 24) CABG ACCESSORY PACK 175619, MODEL/ITEM NUMBER UMCA14E; 25) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42E; 26) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42F; 27) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42G.
- Reason
- During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Classification
- Class II
- Recalling firm
- American Contract Systems Inc
- Quantity
- 31506 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2464-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 1) MODEL/ITEM NUMBER ANCV78BC: UDI/DI 00191072235169, Kit Lot Numbers: 42-7844511, 42-8133811, 42-8037211, 42-7952411, 42-7882011; 2) MODEL/ITEM NUMBER ANVA82AB: UDI/DI 00191072235183, Kit Lot Numbers: 42-8025011, 42-8108511, 42-7928511; 3) MODEL/ITEM NUMBER BHOP48O: UDI/DI 00191072221490, Kit Lot Numbers: 9207467, 8862838, 9140871, 9049946, 9007346, 8934280; 4) MODEL/ITEM NUMBER BHOP48P: UDI/DI 00191072234544, Kit Lot Numbers: 9268782, 9345847; 5) MODEL/ITEM NUMBER BHPH18F: UDI/DI 00191072224651, Kit Lot Numbers: 8993526, 9098058, 8899297; 6) MODEL/ITEM NUMBER BHPH18G: UDI/DI 00191072234551, Kit Lot Numbers: 9238544; 7) MODEL/ITEM NUMBER HGCB98F: UDI/DI 00191072205285, Kit Lot Numbers: 42-7813711; 8) MODEL/ITEM NUMBER HGCV05K: UDI/DI 00191072212474, Kit Lot Numbers: 42-8097611, 42-8021411, 42-7870011, 42-7831911, 42-8093211; 9) MODEL/ITEM NUMBER HGOH99H: UDI/DI 00191072235336, Kit Lot Numbers: 42-8119511, 42-8044311, 42-7874411; 10) MODEL/ITEM NUMBER MHCV10AM: UDI/DI 00191072235619. Kit Lot Numbers: 42-7848211; 11) MODEL/ITEM NUMBER MTVS73D: UDI/DI 00191072237583, Kit Lot Numbers: 12-7397911, 12-7335111, 12-7072111, 12-7142511, 12-7224911; 12) MODEL/ITEM NUMBER RCAV40C: UDI/DI 00191072235381, Kit Lot Numbers: 42-8053511, 42-8151011, 42-7978811, 42-7853911; 13) MODEL/ITEM NUMBER RCOH61G: UDI/DI 00191072229823, Kit Lot Numbers: 42-7803611, 42-7941711, 42-7997711, 42-8075711; 14) MODEL/ITEM NUMBER TNCA08T: UDI/DI 00191072235534, Kit Lot Numbers: 42-7827011, 42-8158811; 15) MODEL/ITEM NUMBER TNCB10AM: UDI/DI 00191072232991, Kit Lot Numbers: 42-7864811, 42-7971311, 42-8065311; 16) MODEL/ITEM NUMBER UDAV16AD: UDI/DI 00191072216229, Kit Lot Numbers: 42-7919811; 17) MODEL/ITEM NUMBER UDCP24AF: UDI/DI 00191072235305, Kit Lot numbers: 42-8010311, 42-8109811; 18) MODEL/ITEM NUMBER UDFP36AH: UDI/DI 00191072235329, Kit Lot Numbers: 42-7835011, 42-7902811, 42-8014011, 42-8091211; 19) MODEL/ITEM NUMBER UICD62AN: UDI/DI 00191072200495, Kit Lot Numbers: 42-7874211, 42-7934511, 42-7999011, 42-8072211, 42-8132211; 20) MODEL/ITEM NUMBER UICS88S: UDI/DI 00191072235428, Kit Lot Numbers: 42-7821511, 42-7913211, 42-7979011, 42-8032611, 42-8106011; 21) MODEL/ITEM NUMBER UIPD03Q: UDI/DI 00191072235435, Kit Lot Numbers: 42-8083011, 42-8154511; 22) MODEL/ITEM NUMBER UITC53AL: UDI/DI 00191072235572, Kit Lot Numbers: 42-7887011, 42-7993911, 42-8071411; 23) MODEL/ITEM NUMBER UIVA75J: UDI/DI 00191072235602, Kit Lot Numbers: 42-7994011, 42-8056011, 42-8114011, 42-7827411, 42-7916411; 24) MODEL/ITEM NUMBER UMCA14E: UDI/DI 00191072222510, Kit Lot Numbers: 334022, 9238536; 25) MODEL/ITEM NUMBER UMOH42E: UDI/DI 00191072222534, Kit Lot Numbers: 8860283; 26) MODEL/ITEM NUMBER UMOH42F: UDI/DI 00191072224712, Kit Lot Numbers: 9125072, 9044687, 8977422, 8925203, 9167965; 27) MODEL/ITEM NUMBER UMOH42G: UDI/DI 00191072234896, Kit Lot Numbers: 9202772, 9334223. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US distribution to states of: IL and LA
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Sep 3, 2025
- Recall initiation date
- Jul 14, 2025
- Center classification date
- Aug 28, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.