Recall archive / FDA device enforcement
Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-... — FDA record Z-2457-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jun 27, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic Sun Stat Japan/600020; Arctic Sun 5000, NA/5000-00-00; Arctic Sun 5000, NA w/ Transmission Interface Module/5000-00-00E; Loaner, Arctic Sun M5000, NA/5000-00-00L; Arctic Sun 5000, EU/5000-1-01; Loaner, Arctic Sun M5000, EU/5000-01-01L; Arctic Sun 5000, UK/5000-01-02; Arctic Sun 5000, Australia/5000-01-03; Arctic Sun 5000, China/5000-01-04; Arctic Sun 5000, Brazil/5000-01-05; Arctic Sun 5000, Switzerland/5000-01-07; Arctic Sun 5000, South Africa/5000-01-08
- Reason
- If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.
- Classification
- Class II
- Recalling firm
- Medivance Inc.
- Quantity
- 10507
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2457-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | REF/UDI-DI/Software: 6000-00-00/00801741161513/v1.0.5 and prior, 6000-00-00L/00801741161520/v1.0.5 and prior, 600001/00801741226380/v2.1.1.0 and prior, 600006/00801741226397/v2.1.1.0 and prior, 600020/00801741226403/v2.1.1.0 and prior, 5000-00-00/00801741080142/v3.0.2 and prior, 5000-00-00E/00801741127755/v3.0.2 and prior, 5000-00-00L/00801741080159/v3.0.2 and prior, 5000-1-01/00801741186134/v3.0.2 and prior, 5000-01-01L/00801741186141/v3.0.2 and prior, 5000-01-02/00801741170003/v3.0.2 and prior, 5000-01-03/00801741222818/v3.0.2 and prior, 5000-01-04/00801741144653/v4.0.0 and prior, 5000-01-05/00801741222825/v3.0.2 and prior, 5000-01-07/00801741222832/v3.0.2 and prior, 5000-01-08/00801741222849/v3.0.2 and prior All serial numbers. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution including in the states of IN, OK, MO, FL, CT, KS, TX, NJ, CA, OR, HI, IL, WI, VA, KY, NH, CO, NV, LA, NC, ID, UT, GA, SC, TN, MD, PA, IA, MA, OH, NY, DE, MI, NM, WV, AZ, DC, VT, MN, SD, AL, AR, WA, NE, ND, MT, AK, MS, ME, WY, RI, PR, OK, OR and the countries of CA, DE, AT, BE, FI, TR, NO, DK, JP, ES, AU, FR, GB, IT, SE, PT, AE, SA, ZA, RU, KR, CY, TH, GR, TW, MY, OM, CN, SG, GI, NL, IL, CZ, RO, CH, RE, PL, PE, SK, AR, IN, CL, MX, QA, BR, SI, HU, JO, LU, KW, PH, NZ, HK, VN, IR, BN, UA, ID, CO, KZ, HR, IS, GU, GE, BH, EE, AM, PA, KR, CL, TH, MY, JP, AU, BD, BE, TW
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Aug 7, 2024
- Recall initiation date
- Jun 27, 2024
- Center classification date
- Jul 31, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.