Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-... — FDA record Z-2457-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 27, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic Sun Stat Japan/600020; Arctic Sun 5000, NA/5000-00-00; Arctic Sun 5000, NA w/ Transmission Interface Module/5000-00-00E; Loaner, Arctic Sun M5000, NA/5000-00-00L; Arctic Sun 5000, EU/5000-1-01; Loaner, Arctic Sun M5000, EU/5000-01-01L; Arctic Sun 5000, UK/5000-01-02; Arctic Sun 5000, Australia/5000-01-03; Arctic Sun 5000, China/5000-01-04; Arctic Sun 5000, Brazil/5000-01-05; Arctic Sun 5000, Switzerland/5000-01-07; Arctic Sun 5000, South Africa/5000-01-08
Reason
If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.
Classification
Class II
Recalling firm
Medivance Inc.
Quantity
10507
Agency-reported status
Ongoing
Source record identifier
Z-2457-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Software: 6000-00-00/00801741161513/v1.0.5 and prior, 6000-00-00L/00801741161520/v1.0.5 and prior, 600001/00801741226380/v2.1.1.0 and prior, 600006/00801741226397/v2.1.1.0 and prior, 600020/00801741226403/v2.1.1.0 and prior, 5000-00-00/00801741080142/v3.0.2 and prior, 5000-00-00E/00801741127755/v3.0.2 and prior, 5000-00-00L/00801741080159/v3.0.2 and prior, 5000-1-01/00801741186134/v3.0.2 and prior, 5000-01-01L/00801741186141/v3.0.2 and prior, 5000-01-02/00801741170003/v3.0.2 and prior, 5000-01-03/00801741222818/v3.0.2 and prior, 5000-01-04/00801741144653/v4.0.0 and prior, 5000-01-05/00801741222825/v3.0.2 and prior, 5000-01-07/00801741222832/v3.0.2 and prior, 5000-01-08/00801741222849/v3.0.2 and prior All serial numbers.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of IN, OK, MO, FL, CT, KS, TX, NJ, CA, OR, HI, IL, WI, VA, KY, NH, CO, NV, LA, NC, ID, UT, GA, SC, TN, MD, PA, IA, MA, OH, NY, DE, MI, NM, WV, AZ, DC, VT, MN, SD, AL, AR, WA, NE, ND, MT, AK, MS, ME, WY, RI, PR, OK, OR and the countries of CA, DE, AT, BE, FI, TR, NO, DK, JP, ES, AU, FR, GB, IT, SE, PT, AE, SA, ZA, RU, KR, CY, TH, GR, TW, MY, OM, CN, SG, GI, NL, IL, CZ, RO, CH, RE, PL, PE, SK, AR, IN, CL, MX, QA, BR, SI, HU, JO, LU, KW, PH, NZ, HK, VN, IR, BN, UA, ID, CO, KZ, HR, IS, GU, GE, BH, EE, AM, PA, KR, CL, TH, MY, JP, AU, BD, BE, TW

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 7, 2024
Recall initiation date
Jun 27, 2024
Center classification date
Jul 31, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.