The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediat... — FDA record Z-2449-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 25, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
Reason
Ground cable installed incorrectly.
Classification
Class II
Recalling firm
Medtronic Navigation, Inc.-Littleton
Quantity
28 units
Agency-reported status
Ongoing
Source record identifier
Z-2449-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel No. BI70002000; UDI-DI: 00763000074128, 00763000081041, 00763000355555, 00763000496784, 00763000542801, 00763000616434, 00763000616526, 00763000616564; Serial No. C2786, C2841, C2781, C2624, C2659, C2660, C2665, C2672, C2708, C2723, C2735, C2760, C2774, C2775, C2779, C2798, C2814, C2823, C2873, C2918, C2968, C3290, C3363, C3188, C3322, C3387, C3404, and C3418.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KY, MI, MO, NC, NV, OH, TN, UT, VA, & WA. The countries of Canada, Ireland, Poland, Russian Federation, Switzerland, & Viet Nam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 30, 2023
Recall initiation date
Jul 25, 2023
Center classification date
Aug 24, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.