DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1... — FDA record Z-2445-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE GLOSS 680 ROLLATOR 1655081 - DOLOMITE GLOSS 600 ROLLATOR 1655082 - DOLOMITE GLOSS 520 ROLLATOR 1655083 - DOLOMITE GLOSS 450 ROLLATOR
Reason
Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user
Classification
Class II
Recalling firm
Invacare Corporation
Quantity
586 devices
Agency-reported status
Ongoing
Source record identifier
Z-2445-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial number in the range: 20BHL0002 to 21IHL1910

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 30, 2023
Recall initiation date
Jun 15, 2023
Center classification date
Aug 23, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.