EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator — FDA record Z-2441-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 11, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator
- Reason
- There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
- Classification
- Class II
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 3,856 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2441-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | GTIN: 00802526544101, 00802526548406, 00802526575105, 00802526575112, 00802526575129, 00802526575136, 00802526575143, 00802526575167, 00802526575181, 00802526575204, 00802526575211, 00802526575228, 00802526599002; All EMBLEM S-ICDs enrolled in LATITUDE |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Aug 30, 2023
- Recall initiation date
- Jul 11, 2023
- Center classification date
- Aug 22, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.