Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product D... — FDA record Z-2433-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 14, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A
Reason
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Classification
Class II
Recalling firm
B Braun Medical Inc
Quantity
14949
Agency-reported status
Ongoing
Source record identifier
Z-2433-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Model No 8713030U; UDI-DI 4046963716745; All Serial Numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide. International distribution to Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 31, 2024
Recall initiation date
Jun 14, 2024
Center classification date
Jul 25, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records