Recall archive / FDA device enforcement
WATCHMAN FXD Curve Access Sys Dbl, OUS, Material Number (UPN) M635TS80020; intended to provide vascular and transsept... — FDA record Z-2430-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 29, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- WATCHMAN FXD Curve Access Sys Dbl, OUS, Material Number (UPN) M635TS80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
- Reason
- Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
- Classification
- Class I
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 340185 units in total
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2430-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | GTIN 00191506013837, ALL NON-EXPIRED BATCHES |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Sep 10, 2025
- Recall initiation date
- Jul 29, 2025
- Center classification date
- Aug 29, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.