Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 77000... — FDA record Z-2420-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 14, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.
Reason
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
Classification
Class II
Recalling firm
MAQUET GMBH
Quantity
315 units (8 US, 307 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-2420-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; UDI-DI: (1) 04046768145177, (2) 04046768145184, (3) 04046768145191, (4) 04046768145207, (5) 04046768145214, (6) 04046768145221; Lot/Serial Numbers: (1) 1, 2, 3, 10, 25, 26, 27, 28, 22, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 24, (2) 9, 51, 53, 14, 15, 16, 17, 34, 35, 36, 37, 39, 40, 42, 43, 54, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 38, 44, 45, 46, 5, 13, 57, 58, 59, 18, 49, 47, 48, 7, 52, (3) 41, 5, 6, 45, 48, 49, 50, 23, 38, 10, 12, 13, 14, 84, 87, 88, 89, 7, 63, 80, 82, 83, 86, 54, 93, 94, 11, 15, 18, 20, 21, 22, 27, 28, 32, 40, 9, 69, 70, 71, 72, 73, 74, 75, 76, 77, 78, 79, 85, 55, 56, 57, 58, 59, 42, 97, 92, 17, 26, 16, 24, 25, 30, 33, 35, 36, 37, 39, 44, 53, 91, 8, 19, 90, 46, 47, 51, 52, 34, 43, 60, 62, 64, 65, 66, 67, 68, (4) 44, 70, 71, 42, 56, 65, 115, 185, 52, 105, 106, 107, 109, 145, 146, 147, 148, 11, 26, 19, 32, 33, 35, 36, 39, 53, 61, 63, 69, 14, 20, 68, 13, 18, 16, 101, 167, 170, 83, 3, 4, 23, 28, 37, 38, 40, 41, 43, 45, 47, 48, 50, 55, 58, 60, 62, 78, 81, 82, 92, 94, 96, 97, 98, 99, 100, 102, 116, 138, 139, 144, 150, 152, 153, 154, 155, 156, 165, 169, 171, 21, 22, 24, 25, 6, 17, 27, 29, 30, 34, 49, 51, 57, 64, 74, 75, 76, 77, 79, 80, 84, 85, 86, 87, 88, 89, 90, 91, 93, 95, 103, 104, 108, 110, 111, 112, 113, 114, 118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, 130, 131, 132, 133, 134, 135, 136, 137, 140, 141, 142, 143, 172, 173, 174, 194, (5) 1, 2, 3, 4, 5, 6, 7, (6) 12, 15;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 3, 2025
Recall initiation date
Jul 14, 2025
Center classification date
Aug 22, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.