Snowden-Pencer MIS DIAMOND-TOUCH CLAMP DEBAKEY 25 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7043; endoscopic surgical device — FDA record Z-2419-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 13, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP DEBAKEY 25 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7043; endoscopic surgical device
Reason
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Classification
Class II
Recalling firm
Carefusion 2200 Inc
Quantity
8 units
Agency-reported status
Ongoing
Source record identifier
Z-2419-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 10885403160547, Lot Numbers: L22, B23, C23

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US and Canada

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Aug 23, 2023
Recall initiation date
Jul 13, 2023
Center classification date
Aug 17, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.