SlideView VS200 slide viewer, — FDA record Z-2416-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 22, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SlideView VS200 slide viewer,
Reason
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
Classification
Class II
Recalling firm
EVIDENT SCIENTIFIC INC
Quantity
N/A
Agency-reported status
Ongoing
Source record identifier
Z-2416-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability65159 67105 67104 69153 67106 69154

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Aug 7, 2024
Recall initiation date
Jul 22, 2024
Center classification date
Jul 26, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.