MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700) — FDA record Z-2397-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
Reason
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Classification
Class II
Recalling firm
Philips North America
Quantity
640 total
Agency-reported status
Ongoing
Source record identifier
Z-2397-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Number: 782120: UDI-DI: 00884838104112; Serial Numbers: 65041 65031 65030 65025 65023 65019 65013 65009 65008 65007 65006 65005 65004 65003 65002 65000 65020 65012 65034 65001; Product Number 782130: UDI-DI: 00884838104402; Serial Numbers: 45301 45298 45007 45297 45557

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide. Global Distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 16, 2023
Recall initiation date
Jun 12, 2023
Center classification date
Aug 10, 2023