UNITRAX Endoprosthesis Head Component - 38mm. Implantable component used in partial hip arthroplasty. Catalog Number:... — FDA record Z-2372-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 19, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
UNITRAX Endoprosthesis Head Component - 38mm. Implantable component used in partial hip arthroplasty. Catalog Number: 6942-5-038
Reason
Mislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
Classification
Class II
Recalling firm
Howmedica Osteonics Corp.
Quantity
12 units
Agency-reported status
Ongoing
Source record identifier
Z-2372-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 07613327032338 Lot Numbers: JT38AD, E94VEV

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, and UK.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 24, 2024
Recall initiation date
Jun 19, 2024
Center classification date
Jul 16, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.