UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a c... — FDA record Z-2364-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 23, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO
Reason
There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
Classification
Class II
Recalling firm
Scholly Fiberoptic Gmbh
Quantity
98 units
Agency-reported status
Ongoing
Source record identifier
Z-2364-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04250480106617 Lot Number: 141980

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide Foreign: Japan, South Korea

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 16, 2023
Recall initiation date
Jun 23, 2023
Center classification date
Aug 9, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.