ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Ma... — FDA record Z-2361-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 19, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000
Reason
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
10 units
Agency-reported status
Ongoing
Source record identifier
Z-2361-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 4056869149325 Serial Numbers: Serial 164840 164851 164854 164853 164855 164856 164833 164836 164845 164850

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 16, 2023
Recall initiation date
Jun 19, 2023
Center classification date
Aug 8, 2023