SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 202... — FDA record Z-2353-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date May 4, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02
- Reason
- Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
- Classification
- Class II
- Recalling firm
- Micromed S.p.A.
- Quantity
- 207 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2353-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of FL, PA.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 17, 2026
- Recall initiation date
- May 4, 2026
- Center classification date
- Jun 5, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.