LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-Indicated for patients with Mobility-limiting diseases, f... — FDA record Z-2345-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 3, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 14-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur. Item Number: 136-114/00
Reason
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
Classification
Class II
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
14 units
Agency-reported status
Ongoing
Source record identifier
Z-2345-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04026575182084 Lot Numbers: B923005 B932048

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 16, 2023
Recall initiation date
Jul 3, 2023
Center classification date
Aug 4, 2023