GE HealthCare OEC 9900 Elite C-arm Systems. — FDA record Z-2342-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 2, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
GE HealthCare OEC 9900 Elite C-arm Systems.
Reason
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Classification
Class II
Recalling firm
GE Medical Systems, LLC
Quantity
Not supplied
Agency-reported status
Ongoing
Source record identifier
Z-2342-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 00840682114349

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 3, 2025
Recall initiation date
Jul 2, 2025
Center classification date
Aug 22, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.