The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended L... — FDA record Z-2337-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date May 31, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.
- Reason
- There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
- Classification
- Class II
- Recalling firm
- Maquet Cardiovascular, LLC
- Quantity
- 600 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2337-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 00607567700345 Lot Numbers; 3000371260 3000378288 3000362934 3000371798 3000378554 3000372477 3000378950 3000376542 3000380314 3000377320 3000380847 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
NationwideForeign: AT DE DK ES GB HK IT NL ZA
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Jul 17, 2024
- Recall initiation date
- May 31, 2024
- Center classification date
- Jul 11, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.