ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. ... — FDA record Z-2325-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 24, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187
Reason
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Classification
Class I
Recalling firm
Boston Scientific Corporation
Quantity
247,702 US; 344,670 OUS
Agency-reported status
Ongoing
Source record identifier
Z-2325-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel 0161 UDI-DI/GTIN 00802526372766; Model 0164 UDI-DI/GTIN N/A; Model 0165 UDI-DI/GTIN 00802526377259; Model 0167 UDI-DI/GTIN N/A 00802526373251; Model 0170 UDI-DI/GTIN 00802526373282 00802526373299; Model 0171 UDI-DI/GTIN 00802526373374 00802526373381; Model 0172 UDI-DI/GTIN 00802526373466; Model 0173 UDI-DI/GTIN 00802526373558 00802526373565; Model 0174 UDI-DI/GTIN 00802526373640; Model 0175 UDI-DI/GTIN 00802526373732 00802526373749; Model 0176 UDI-DI/GTIN 00802526376726 00802526376733; Model 0177 UDI-DI/GTIN 00802526376818 00802526376825; Model 0180 UDI-DI/GTIN 00802526411977 00802526422027 00802526433597 00802526531149 00802526540943 00802526540950 00802526591518; Model 0181 UDI-DI/GTIN 00802526411984 00802526422034 00802526433689 00802526433726 00802526433757 00802526491009 00802526502903 00802526531156; Model 0182 UDI-DI/GTIN 00802526422041 00802526531163; Model 0183 UDI-DI/GTIN 00802526422058 00802526531170; Model 0184 UDI-DI/GTIN 00802526412011 00802526422065 00802526433986 00802526433993 00802526491313 00802526531187 00802526541261 00802526583438; Model 0185 UDI-DI/GTIN 00802526412028 00802526413438 00802526422072 00802526434082 00802526434099 00802526434105 00802526434150 00802526434174 00802526470622 00802526491405 00802526491412 00802526491429 00802526503306 00802526531194 00802526541346 00802526591563 00802526612541; Model 0186 UDI-DI/GTIN 00802526412035 00802526422089 00802526434198 00802526531200; Model 0187 UDI-DI/GTIN 00802526412042 00802526422096 00802526531217

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution. US nationwide. International distribution worldwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 27, 2025
Recall initiation date
Jul 24, 2025
Center classification date
Aug 20, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.