Recall archive / FDA device enforcement
ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. ... — FDA record Z-2325-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jul 24, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187
- Reason
- Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
- Classification
- Class I
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 247,702 US; 344,670 OUS
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2325-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model 0161 UDI-DI/GTIN 00802526372766; Model 0164 UDI-DI/GTIN N/A; Model 0165 UDI-DI/GTIN 00802526377259; Model 0167 UDI-DI/GTIN N/A 00802526373251; Model 0170 UDI-DI/GTIN 00802526373282 00802526373299; Model 0171 UDI-DI/GTIN 00802526373374 00802526373381; Model 0172 UDI-DI/GTIN 00802526373466; Model 0173 UDI-DI/GTIN 00802526373558 00802526373565; Model 0174 UDI-DI/GTIN 00802526373640; Model 0175 UDI-DI/GTIN 00802526373732 00802526373749; Model 0176 UDI-DI/GTIN 00802526376726 00802526376733; Model 0177 UDI-DI/GTIN 00802526376818 00802526376825; Model 0180 UDI-DI/GTIN 00802526411977 00802526422027 00802526433597 00802526531149 00802526540943 00802526540950 00802526591518; Model 0181 UDI-DI/GTIN 00802526411984 00802526422034 00802526433689 00802526433726 00802526433757 00802526491009 00802526502903 00802526531156; Model 0182 UDI-DI/GTIN 00802526422041 00802526531163; Model 0183 UDI-DI/GTIN 00802526422058 00802526531170; Model 0184 UDI-DI/GTIN 00802526412011 00802526422065 00802526433986 00802526433993 00802526491313 00802526531187 00802526541261 00802526583438; Model 0185 UDI-DI/GTIN 00802526412028 00802526413438 00802526422072 00802526434082 00802526434099 00802526434105 00802526434150 00802526434174 00802526470622 00802526491405 00802526491412 00802526491429 00802526503306 00802526531194 00802526541346 00802526591563 00802526612541; Model 0186 UDI-DI/GTIN 00802526412035 00802526422089 00802526434198 00802526531200; Model 0187 UDI-DI/GTIN 00802526412042 00802526422096 00802526531217 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide Distribution. US nationwide. International distribution worldwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Aug 27, 2025
- Recall initiation date
- Jul 24, 2025
- Center classification date
- Aug 20, 2025