Volara System. Single Patient Use (SPU) Circuit 5kit for Volara and Single Patient Use (SPU) Circuit kit for Volara. ... — FDA record Z-2321-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 30, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Volara System. Single Patient Use (SPU) Circuit 5kit for Volara and Single Patient Use (SPU) Circuit kit for Volara. Intermittent positive pressure breathing device.
Reason
Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
Classification
Class I
Recalling firm
Baxter Healthcare Corporation
Quantity
58,220 units
Agency-reported status
Ongoing
Source record identifier
Z-2321-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Codes: M08473/M08474; Lot numbers: 2171WA, 2237WA, 2238WA, 2241WA, 2242WA, 2243WA, 2244WA, 2249WA, 2250WA, 2251WA, 2285WA, 2286WA, 2290WA, 2297WA, 2298WA, 2299WA, 2307WA, 2308WA, 2311WA, 2332WA, 2333WA, 2341WA, 2342WA, 2349WA, 2355WA, 2356WA, 2357WA, 2363WA, 2364WA, 3019WA, 3061WA, 3067WA, 3068WA, 3069WA, 3072WA, 3073WA, 3074WA, 3080WA, 3081WA, 3094WA, 3095WA, 3096WA, 3100WA, 3101WA, 3102WA, 3103WA, 3109WA, 3110WA, 3111WA, 3123WA, 3124WA, 3152WA, 3156WA, 3157WA, 3158WA, 3213WA, 3214WA, 3215WA, 3219WA, 3240WA, 3241WA, 3268WA, 3270WA, 3271WA, 3272WA, 3297WA, 3298WA, 3299WA, 3300WA, 3303WA, 3304WA, 3305WA, 3306WA, 3310WA, 3311WA, 3313WA, 3326WA, 3327WA, 3328WA, 3331WA.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide distribution. International distribution to Canada, France, and Germany.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 24, 2024
Recall initiation date
May 30, 2024
Center classification date
Jul 17, 2024