Recall archive / FDA device enforcement
Allia IGS 7 Pulse angiographic X-ray system — FDA record Z-2316-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Apr 20, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Allia IGS 7 Pulse angiographic X-ray system
- Reason
- GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
- Classification
- Class II
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 35 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2316-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 00195278719720; Serial Numbers: D3-23-024, D3-23-029, D3-24-001, D3-24-002, D3-24-005, D3-24-006, D3-24-008, D3-24-009, D3-24-010, D3-24-013, D3-24-014, D3-24-016, D3-24-019, D3-24-020, D3-24-021, D3-24-023, D3-24-025, D3-25-001, D3-25-002, D3-25-003, D3-25-005, D3-25-006, D3-25-007, D3-25-008, D3-25-009, D3-25-010, D3-25-013, D3-25-016, D3-25-017, D3-25-018, D3-25-026, D3-25-027, D3-25-029, D3-25-030, D3-25-034, D3-25-036, D3-25-037, D3-26-001, D3-26-002, D3-26-007 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 17, 2026
- Recall initiation date
- Apr 20, 2026
- Center classification date
- Jun 5, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.