Plum Solo Precision IV Pump, 40001-0401 — FDA record Z-2311-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 29, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Plum Solo Precision IV Pump, 40001-0401
Reason
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Classification
Class II
Recalling firm
ICU Medical, Inc.
Quantity
13,613
Agency-reported status
Ongoing
Source record identifier
Z-2311-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI M335400011 Plum Solo Precision IV Pump v1.0.3

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 17, 2026
Recall initiation date
Apr 29, 2026
Center classification date
Jun 5, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.