PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050 — FDA record Z-2275-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 27, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
Reason
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Classification
Class II
Recalling firm
Oculus Technologies of Mexico, S.A. de C.V.
Quantity
5904 units
Agency-reported status
Ongoing
Source record identifier
Z-2275-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI:00860009268616 Lot: 25G263

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 10, 2026
Recall initiation date
Apr 27, 2026
Center classification date
Jun 1, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.