Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tub... — FDA record Z-2262-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 22, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.
Reason
Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
Classification
Class II
Recalling firm
Baxter Healthcare Corporation
Quantity
105675
Agency-reported status
Ongoing
Source record identifier
Z-2262-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 10, 2026
Recall initiation date
Apr 22, 2026
Center classification date
May 29, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records