Recall archive / FDA device enforcement
Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hem... — FDA record Z-2258-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Apr 22, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).
- Reason
- Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.
- Classification
- Class II
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Quantity
- 145 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2258-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI N/A for the Prucka 3 Amplifier-not a medical device. CardioLab and ComboLab Systems GTIN #00195278507044 and #00195278507051. Prucka 3 Amplifier serial numbers: SVS23400020SA, SVS23400024SA, SVS23360017SA, SVS23400003SA, SVS23470003SA, SVS23350005SA, SVS23330008SA, SVS23350001SA, SVS23350003SA, SVS23360018SA, SVS23520022SA, SVS23350008SA, SVS23280009SA, SVS23350021SA, SVS23470016SA, SVS23470010SA, SVS23360013SA, SVS23280002SA, SVS23400026SA, SVS23400039SA, SVS23400021SA, SVS23330006SA, SVS23360014SA, SVS23360016SA, SVS23460020SA, SVS23470022SA, SVS23480007SA, SVS23360010SA, SVS23350024SA, SVS24020002SA, SVS24020004SA, SVS24020005SA, SVS24020006SA, SVS24020007SA, SVS24020011SA, SVS23460002SA, SVS23350004SA, SVS23350006SA, SVS23350007SA, SVS23480003SA, SVS23400008SA, SVS23210005SA, SVS23360007SA, SVS23360002SA, SVS23360012SA, SVS23470004SA, SVS23520018SA, SVS23520020SA, SVS24020016SA, SVS23460001SA, SVS23470006SA, SVS23470008SA, SVS23350002SA, SVS23350009SA, SVS23350025SA, SVS23360003SA, SVS23360005SA, SVS23360006SA, SVS23360011SA, SVS23400023SA, SVS23400030SA, SVS23400031SA, SVS23400032SA, SVS23400037SA, SVS23400038SA, SVS23460007SA, SVS23400015SA, SVS23350011SA, SVS23350017SA, SVS23470017SA, SVS23470018SA, SVS23470020SA, SVS23460013SA, SVS23460014SA, SVS23480002SA, SVS23480005SA, SVS23480010SA, SVS23480011SA, SVS23480013SA, SVS23480015SA, SVS23520012SA, SVS23280008SA, SVS23290004SA, SVS23330004SA, SVS23350014SA, SVS23350022SA, SVS23350023SA, SVS23460006SA, SVS23400022SA, SVS23350012SA, SVS23350016SA, SVS23480004SA, SVS23470007SA, SVS23400025SA, SVS23460003SA, SVS23460004SA, SVS23350015SA, SVS23320002SA, SVS23470013SA, SVS23350020SA, SVS23330005SA, SVS23470009SA, SVS23470011SA, SVS23470019SA, SVS23470012SA, SVS23460016SA, SVS23290001SA, SVS23350013SA, SVS23460017SA, SVS23480012SA, SVS23480016SA, SVS23330007SA, SVS23360008SA, SVS23360020SA, SVS23400005SA, SVS23400009SA, SVS23400010SA, SVS23400019SA, SVS23460008SA, SVS23280005SA, SVS23400004SA, SVS23460005SA, SVS23400034SA, SVS23400035SA, SVS23400036SA, SVS23400033SA, SVS23400018SA, SVS23470005SA, SVS23400028SA, SVS23400029SA, SVS23360019SA, SVS23400002SA, SVS23360004SA, SVS23400007SA, SVS23400011SA, SVS23400006SA, SVS23350018SA, SVS23330003SA, SVS23290003SA, SVS23290002SA, SVS23330009SA, SVS23350019SA, SVS23360015SA, SVS23470001SA, and SVS23470002SA. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL, GA, IL, KY, MA, ME, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. The countries of Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Japan, Republic of Korea, and Singapore.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jul 10, 2024
- Recall initiation date
- Apr 22, 2024
- Center classification date
- Jul 3, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.