NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 ... — FDA record Z-2255-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 24, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems (NeuMoDx System(s) Ref: 100100
Reason
Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results
Classification
Class II
Recalling firm
NeuMoDx Molecular Inc
Quantity
20,016 cartridges
Agency-reported status
Ongoing
Source record identifier
Z-2255-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 10814278020274 Lot Numbers: 117514, 117524, and 117528

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution to states of: Kansas, Florida, Texas, Ohio, New York, Michigan, Pennsylvania; and OUS (Foreign) to countries of: AT, BE, CH, DE, ES, FI, FR, GB, IT, LV, PL, RO, SE

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 9, 2023
Recall initiation date
May 24, 2023
Center classification date
Jul 28, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records