Carina Sub-Acute Care Ventilator — FDA record Z-2248-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Carina Sub-Acute Care Ventilator
Reason
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Classification
Class I
Recalling firm
Draeger Medical, Inc.
Quantity
11,621 units
Agency-reported status
Ongoing
Source record identifier
Z-2248-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPart No. 5704110; UDI-DI 04048675398516; All Serial No.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 30, 2023
Recall initiation date
Jul 12, 2023
Center classification date
Aug 18, 2023