Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensi... — FDA record Z-2240-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 5, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088
Reason
Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
9 units
Agency-reported status
Ongoing
Source record identifier
Z-2240-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 2, 2023
Recall initiation date
Jun 5, 2023
Center classification date
Jul 27, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.