Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. I... — FDA record Z-2237-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 29, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. It is an in-vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids with Catalog Number C11137.
Reason
The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000 Access Immunoassay Analyzer may generate a System Event log message that indicates PnP errors. Any cancelled tests will have an associated SYS flag. The analyzer may enter the Red-system status because of this issue. The analyzer also may not eject RVs, where unbound conjugate could potentially splash up on the sidewall of the RV, not being properly washed away, resulting in an artificially higher signal (RLU). The analyzer may not detect this splashing and continue with normal operation, which can cause false low or high erroneous results. Consequently, the analyzer may cancel in-progress tests. These errors could potentially cause delay reporting patient test results and patient treatments. The degree of the potential health hazard is dependent on the analytes that are potentially delayed, especially those analytes that would be used in the acute setting when treatment decisions are immediate. In this particular recall the highest risk analyte is troponin. Health consequences could include permanent impairment or possibly death if medical treatment decisions are delayed due to a delay in obtaining patient results. The issue was identified by Beckman Coulter Inc. manufacturing staff at their location in Brea California on February 29, 2024, during in-process acceptance inspection activities.
Classification
Class II
Recalling firm
Beckman Coulter, Inc.
Quantity
9 instruments
Agency-reported status
Ongoing
Source record identifier
Z-2237-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 15099590732103, Serial Numbers: 300273, 300275, 300276, 300277, 300281, 300282, 300284, 300286, 300297

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the state of Indiana and the countries of Italy, Sweden, Ireland, Spain.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 10, 2024
Recall initiation date
Feb 29, 2024
Center classification date
Jun 28, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.