What the record says

Product
Ringed DxI Reaction Vessels (RVs)
Reason
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
Classification
Class II
Recalling firm
Beckman Coulter, Inc.
Quantity
213 cases of 10,000 RVs (2,130,000 RVs)
Agency-reported status
Ongoing
Source record identifier
Z-2237-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 15099590229979; Catalog Number: 386167; Lot Numbers: L17855111, L17798598, L18046334

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 2, 2023
Recall initiation date
Jun 20, 2023
Center classification date
Jul 26, 2023