Product
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Reason
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Classification
Class I
Recalling firm
Fresenius Kabi USA, LLC
Quantity
18,444
Agency-reported status
Ongoing
Source record identifier
Z-2224-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;