Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700,... — FDA record Z-2221-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 13, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.
Reason
Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"
Classification
Class II
Recalling firm
Volcano Corp
Quantity
233,817
Agency-reported status
Ongoing
Source record identifier
Z-2221-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00845225002848, 00184360000037. Instructions for use: 300004875732_A Revision Date: 03/2021, 300004410212.A Release Date: 03/2021

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of BR, PL, SK, IT, DK, DE, GB, AT, KW, CR, ES, NL, LV, MT, IE, CH, ZA, RO, FI, PR, FR, CA, PA, IL, SE, PT, HR, GR, NZ, HK, TW, MY, TH, MX, AR, PE, AE, BG, AU, JP, ID, RS, CO, SG, BE.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 6, 2025
Recall initiation date
Jun 13, 2025
Center classification date
Jul 30, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records