SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in... — FDA record Z-2221-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 23, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES D684-00-0568-01 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) D684-00-0568-03 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE D684-00-0568-05 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) D684-00-0568-09 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY D684-00-0568-01U PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB D884-00-0019-22
Reason
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Classification
Class II
Recalling firm
Datascope Corp.
Quantity
321,609 total kits
Agency-reported status
Ongoing
Source record identifier
Z-2221-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Description UDI SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES 10607567108063 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) 10607567109428 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE 10607567109442 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) 10607567113166 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY 10607567109565 PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB 10607567108612

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide. International distribution worldwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 2, 2023
Recall initiation date
Jun 23, 2023
Center classification date
Jul 21, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records