Recall archive / FDA device enforcement
TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta ... — FDA record Z-2218-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jun 23, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D884-00-0019-18 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs D884-00-0019-26
- Reason
- During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
- Classification
- Class II
- Recalling firm
- Datascope Corp.
- Quantity
- 321,609 total kits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2218-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Product Description UDI 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES 10607567107656 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES 10607567107663 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES 10607567111711 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES 10607567111728 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs 10607567107684 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs 10607567111810 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Domestic distribution nationwide. International distribution worldwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Aug 2, 2023
- Recall initiation date
- Jun 23, 2023
- Center classification date
- Jul 21, 2023