Pediatric care bed; Product Designation: KayserBett IDA; — FDA record Z-2212-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 4, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Pediatric care bed; Product Designation: KayserBett IDA;
Reason
If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).
Classification
Class I
Recalling firm
KAYSERBETTEN GMBH & CO. KG
Quantity
372 units
Agency-reported status
Ongoing
Source record identifier
Z-2212-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6338, 6339, 6340, 6341, 6342, 6343, 6344, 6345, 6346, 6347, 6351, 6355, 6795, 6796, 7176, 7177, 7178, 7179, 7180, 7181, 7182, 7183, 7184, 7185, 7186, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7617, 7618, 7619, 7620, 7621, 7622, 7623, 8242, 8243, 8244, 8245, 8588, 8589, 8590, 8591, 8592, 8593, 8594, 8595, 8596, 8597, 8598, 8599, 8600, 8601, 8602, 8603, 8608, 8609, 8691, 8704, 8912, 8921, 8922, 8923, 8924, 8925, 8926, 8927, 8928, 8929, 8930, 9725, 9726, 9727, 10350, 10351, 10352, 10819, 10820, 10821, 10822, 10823, 10824, 10825, 11186, 11367, 11368, 11369, 11370, 11371, 11372, 11797, 11798, 11799, 11800, 11801, 11802, 11803, 11804, 11805, 11806, 11807, 11808, 12247, 12248, 12249, 12250, 12251, 12252, 12253, 12254, 12255, 12256, 12257, 12397, 12398, 12399, 12782, 12783, 12963, 12964, 12965, 12966, 12967, 12968, 12969, 12970, 12971, 12972, 13486, 13487, 13488, 13489, 13490, 13491, 13492, 13493, 13494, 13495, 13496, 13497, 13912, 13913, 13914, 13915, 13916, 13917, 13918, 14942, 14944, 14952, 14953, 15912, 15913, 15914, 16403, 16404, 16405, 16406, 16407, 16408, 16409, 16410, 16411, 16984, 16985, 16986, 16987, 16988, 16989, 16990, 16991, 16992, 16993, 16994, 17577, 17578, 17579, 17580, 17581, 17582, 17583, 17584, 17585, 17586, 17587, 18494, 19112, 19113, 19114, 20151, 20152, 20153, 20154, 20155, 20156, 20157, 20158, 20963, 20964, 20965, 20966, 20967, 20968, 20969, 20970, 20972, 21192, 21193, 21194;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distribution US Nationwide and Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 10, 2026
Recall initiation date
May 4, 2026
Center classification date
Jun 4, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.