Recall archive / FDA device enforcement
Automated Impella Controller (AIC); Product Code: 0042-0000-US; — FDA record Z-2211-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Apr 20, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Automated Impella Controller (AIC); Product Code: 0042-0000-US;
- Reason
- A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.
- Classification
- Class I
- Recalling firm
- Abiomed, Inc.
- Quantity
- 91 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2211-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783; |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jun 3, 2026
- Recall initiation date
- Apr 20, 2026
- Center classification date
- May 22, 2026