Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoL... — FDA record Z-2210-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 24, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960
Reason
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential for the patient to receive less infusion solution than intended. Exposure of hazardous substances to the patient, caregiver, or pharmacy technician is possible.
Classification
Class II
Recalling firm
ICU Medical, Inc.
Quantity
20,506
Agency-reported status
Ongoing
Source record identifier
Z-2210-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 10887709083278. Lots: 13919794, 14134777, 14196319

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX, FL, AR, CT, PA, WA, WI, CO, VT, NY, NH, KY, IA, KS, AL, NM, IL, MO, NJ.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 27, 2026
Recall initiation date
Mar 24, 2026
Center classification date
May 18, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.