BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60... — FDA record Z-2180-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 8, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV; REF: 2452-0007, BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 10D 3SS 2CV; REF: 2204-0007, BD Alaris Pump Infusion SetAS LVP 20D; REF: 24001-0007, BD Alaris Pump Infusion Set, Back Check Valve, SmartSite Y-siteAS LVP 20D SS CV; REF: 10015012, BD Alaris Pump Infusion Burette Set, 0.2 Micron Filter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-siteAS LVP BUR 20D PE-Lined SMBORE SS 0.2M; REF: 2432-0007, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D 3SS 0.2M CV; REF: 2434-0007, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, SmartSite Y-siteAS LVP 20D SS 0.2M CV; REF: 2202-0007, BD Alaris Pump Infusion Set, 1.2 Micron FilterAS LVP 20D 1.2M; REF: 11426965, BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sitesAS LVP 20D CV; REF: 10561554, BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D DEHP FREE 3SS 15M CV; REF: 10863358, BD Alaris Pump Infusion Set, 15 Micron FilterAS LVP 20D DEHP FREE 15M; REF: 11171447, BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 20D 3SS 2CV; REF: 11532269, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sitesAS LVP 20D PE-LINED 2SS 0.2M CV; REF: 10010454, BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-siteAS LVP 20D PE-LINED TUBING SS 0.2M; REF: 2465-0007, BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, Low Sorbing Tubing (PE Lined), SmartSite Y-siteAS LVP SS BAG 20D PE-Lined SS 0.2M; REF: 10010453, ASBD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site LVP 20D PE-LINED TUBING SS 1.2M REF: 11419365, BD Alaris" Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 SmartSite" Y-sites REF: 10813621, BD Alaris" Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 2423-0007, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 10015896, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 3 SmartSite" Y-sites REF: 24010-0007T, BD Alaris" Pump Infusion Set, Bonded Texium", Closed Male Luer with Priming Cap, Back Check Valve, 3 SmartSite" Y-sites REF: 10010483, BD Alaris" Pump Infusion Set, Vented Syringe Adapter, Smallbore Tubing REF: 2403-0007, BD Alaris" Pump Infusion Set, Half Set, SmartSite" Y-site REF: 24301-0007T, BD Alaris" Pump Infusion Set, SmartSite" Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, Bonded Texium" Closed Male Luer with Priming Cap, SmartSite" Y-site
Reason
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Classification
Class I
Recalling firm
CareFusion 303, Inc.
Quantity
33,162,680 units total
Agency-reported status
Ongoing
Source record identifier
Z-2180-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF: 10013186/ UDI: 07613203021173; REF: 2452-0007/ UDI: 10885403219870; REF: 2204-0007/ UDI: 10885403199363; REF: 24001-0007/ UDI: 10885403238666; REF: 10015012/ UDI: 10885403233951; REF: 2432-0007/ UDI: 10885403232329; REF: 2434-0007/ UDI: 07613203019682; REF: 2202-0007/ UDI: 10885403274039; REF: 11426965/ UDI: 10885403232558; REF: 10561554/ UDI: 10885403232565; REF: 10863358/ UDI: 10885403232466; REF: 11171447/ UDI: 07613203021234; REF: 11532269/ UDI: 10885403232343; REF: 10010454/ UDI: 07613203015806; REF: 2465-0007/ UDI: 10885403221941; REF: 10010453/ UDI: 07613203021135 REF: 11419365/ UDI: 10885403234736 REF: 10813621/UDI: 10885403234514 REF: 2423-0007/UDI: 07613203020985 REF: 10015896/UDI: 10885403234064 REF: 24010-0007T/UDI: 10885403271021 REF: 10010483/UDI: 10885403232602 REF: 2403-0007/UDI: 10885403232473 REF: 24301-0007T/UDI: 10885403223198

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 13, 2025
Recall initiation date
Jul 8, 2025
Center classification date
Aug 6, 2025