- Product
- Titan Blis-Sonic Scaler packaged with the Titan Scaler Tip - Perio (part number 261669) (1) P/N 264563 (2) P/N 264565 The scaler tip is used with an air drive scaler for subgingival scaling and fine access. The Perio tip is used for lingual supragingival scaling of mandibular incisors and deep local pockets.
- Reason
- Tip holder portion of the scaler tip assembly may have an oversized condition that will not allow the scaler tip to stay seated in the holder may cause injury
- Classification
- Class II
- Recalling firm
- Dental EZ Group Star Dental Division
- Quantity
- 192 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2178-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI : D7022645630 (1) P/N 264563: lot # 4874201, lot # 4904771 (2) P/N 264565, lot # 4964531 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jul 3, 2024
- Recall initiation date
- May 30, 2024
- Center classification date
- Jun 24, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.