Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: ... — FDA record Z-2163-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 14, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System
Reason
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
Classification
Class II
Recalling firm
Dexcom, Inc.
Quantity
58582
Agency-reported status
Ongoing
Source record identifier
Z-2163-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: SW12300 versions 2.9.0 through 2.11.2 - UDI-DI: 00386270001863

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2: Worldwide - US Nationwide including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Austria, Australia, Belgium, Bahrain, Canada, Czech Republic, Germany, Denmark, Estonia, Finland, France, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, Kuwait, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom. SW14245 Dexcom ONE iOS App versions 1.5.0 and 1.6.0 - Only International distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Czech Republic, Estonia, France, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Malaysia, Monaco, Netherlands, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Spain, Sweden, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 20, 2026
Recall initiation date
Apr 14, 2026
Center classification date
May 11, 2026