Simplexa Direct Amplification Disc Kit, Rx Only, IVD The Simplexa Direct Amplification Disc Kit, REF: MOL1455 is 1 bo... — FDA record Z-2153-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 14, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Simplexa Direct Amplification Disc Kit, Rx Only, IVD The Simplexa Direct Amplification Disc Kit, REF: MOL1455 is 1 box containing 3 Direct Amplification Discs (DADs). Each box contains discs individually pouch-sealed and labeled with REF: MOL1452
Reason
There is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc.
Classification
Class II
Recalling firm
DiaSorin Molecular LLC
Quantity
4,7878 units
Agency-reported status
Ongoing
Source record identifier
Z-2153-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityMOL1455 UDI-DI: 01-30816101025092 MOL1455 kit lot codes: P17044N P17052N P17301N P17302N P17303N P17304N P17305N P17306N P17351N P17352N P17353N P17354N P17614N P17625N P17650N P17652N P17653N P17654N P17687N P17688N P17731N MOL1452 disc lot codes: P15886N P16794N P16795N P16796N P16883N P16885N P17022N P17056N P17057N P17047N P17023N P17176N P17177N P17294N P17178N P17295N P17359N P17362N P17296N P17361N P17046N

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SD, TN, TX, VA, VT, WA, WI, WV and the countries of Bangladesh, Canada, Israel, Italy.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 26, 2023
Recall initiation date
Jun 14, 2023
Center classification date
Jul 18, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.