Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457 — FDA record Z-2128-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 26, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Reason
There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach.
Classification
Class II
Recalling firm
Abiomed, Inc.
Quantity
37 units
Agency-reported status
Ongoing
Source record identifier
Z-2128-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00813502013252; Lot Numbers: 1727033, 1725590

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of MN, NJ, TN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 19, 2023
Recall initiation date
May 26, 2023
Center classification date
Jul 11, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records