Recall archive / FDA device enforcement
Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501... — FDA record Z-2127-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jan 7, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069
- Reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Classification
- Class II
- Recalling firm
- Medline Industries, LP
- Quantity
- 175 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-2127-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 1) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 21JBJ732; 2) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22ABJ831; 3) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22BBD031; 4) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22FBP788; 5) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22IBE795; 6) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22IBF936; 7) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 23BBJ847; 8) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 23BBL738; 9) DYNJ22501G, UDI-DI: 10889942234110(each), 40889942234111(case), Lot Number: 21ABK153; 10) DYNJ22501G, UDI-DI: 10889942234110(each), 40889942234111(case), Lot Number: 21CBZ902; 11) DYNJ32715F, UDI-DI: 10889942530038(each), 40889942530039(case), Lot Number: 21ADB529; 12) DYNJ50834C, UDI-DI: 10889942568178(each), 40889942568179(case), Lot Number: 21BDA710; 13) DYNJ51069, UDI-DI: 10889942118410(each), 40889942118411(case), Lot Number: 21BDB116 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- May 13, 2026
- Recall initiation date
- Jan 7, 2026
- Center classification date
- May 7, 2026